SportsMed ReviewFor Providers
Resources

PRP Kit Reference Guide

A reference to the major platelet-rich plasma systems used in orthopedic surgery and non-operative sports medicine. Filter by leukocyte type, spin protocol, or brand to match the right PRP preparation to your clinical indication.

32 systems

Arthrex

3 products
ACP (Autologous Conditioned Plasma)
LP-PRP
Spin
Single Spin
Concentration
2–3× baseline
Output
~4–7 mL from 15 mL blood
Centrifuge
Hettich Rotofix 32 (Arthrex-branded swinging bucket). Single spin, ~5 min.

The most commonly used office-based PRP in sports medicine. Leukocyte- and RBC-depleted, low-volume preparation. Strong evidence base for knee OA, hip OA, lateral epicondylitis, and some tendinopathies. Quick prep (~5 min spin) and minimal post-injection inflammation.

ACP Max
LP-PRP
Spin
Double Spin
Concentration
Up to 12× baseline
Output
From 30 / 60 / 90 mL blood draws
Centrifuge
Drucker Horizon 24 Flex (Arthrex-paired). Double spin.

A higher-dose, double-spin evolution of ACP. Produces leukocyte- and neutrophil-depleted PRP at significantly higher platelet concentrations than standard ACP. Indicated when a larger therapeutic platelet dose is desired (knee OA, advanced tendinopathy, surgical adjunct).

Angel System
Customizable (LR or LP)
Spin
Automated 3-Sensor (flow cytometry)
Concentration
Up to 18× baseline
Output
Customizable; processes up to 180 mL blood
Centrifuge
Arthrex Angel device — proprietary, fully automated 3-sensor optical separation.

Fully automated system using 3-sensor optical separation. Adjustable hematocrit and leukocyte settings let the operator dial in LR-PRP for tendinopathies or LP-PRP for intra-articular use, from the same disposable. Common in OR settings (cartilage repair, rotator cuff, microfracture adjunct) and high-volume regenerative clinics.

EmCyte

4 products
PurePRP SP (60 mL / 120 mL)
Customizable (LR or LP)
Spin
Double Spin (Supraphysiologic)
Concentration
Up to 15×+ baseline
Output
~7 mL from up to 104 mL blood
Centrifuge
EmCyte Executive Series II (Eppendorf-based, manual, max ~3000g) or Platinum/Sapphire Series (Elmi-based, preset). Double spin: 1.5 min @ 3800 RPM, then 1–4 min @ 4400 RPM.

EmCyte's flagship double-spin platform. Uniquely enriches monocytes while letting the operator control neutrophil inclusion — making it useful across both tendinopathies (LR setting) and intra-articular OA (LP setting). Strong adoption in non-operative regenerative sports medicine.

PurePRP II
Customizable (LR or LP)
Spin
Double Spin
Concentration
~6–10× baseline
Output
~6–10 mL PRP from 60 mL blood
Centrifuge
EmCyte Executive II / Platinum / Sapphire Series. Double spin, ~8 min total.

Earlier-generation double-spin PurePRP kit, still widely used. Same leukocyte selectivity principle as PurePRP SP at a lower-volume processing footprint. Workhorse for knee/hip OA and tendinopathy in office-based regenerative practices.

AbsolutePRP (GOLD Series)
LP-PRP
Spin
Double Spin
Concentration
Variable
Output
~4–6 mL PRP from 20 mL blood (small-volume kit)
Centrifuge
EmCyte Executive II / Platinum / Sapphire Series. Single spin.

Lower-volume kit in the EmCyte line. Useful for small-joint, facet, or peritendinous injections where lower total volume is needed.

PureBMC SP
BMAC + PRP
Spin
Double Spin
Concentration
High concentrate of MNCs + platelets
Output
~7 mL BMC from 60–120 mL bone marrow aspirate
Centrifuge
EmCyte Executive II / Platinum / Sapphire Series.

Bone marrow aspirate concentrate (BMAC) system that produces a cell concentrate alongside PRP. Indicated for advanced cartilage lesions, AVN, non-union, and recalcitrant tendinopathies.

Zimmer Biomet

2 products
GPS III
LR-PRP
Spin
Single Spin (buoy-based)
Concentration
~9.3× baseline (â‰Ĩ5× WBC)
Output
~6 mL from 60 mL blood
Centrifuge
Zimmer Biomet proprietary centrifuge. Single spin: 15 min @ 3200 RPM.

A long-standing leukocyte-rich PRP system widely used in orthopedic ORs and hospital systems. â‰Ĩ90% platelet recovery with elevated WBC concentration. Used surgically (mixed with autograft/allograft bone) and clinically for tendinopathy. Strong literature base — frequently used as the comparator in PRP studies.

Plasmax
PPP Concentrator (Adjunct)
Spin
Polyacrylamide bead concentration
Concentration
Fibrinogen-rich PPP concentrate
Output
Centrifuge
Zimmer Biomet centrifuge (same as GPS III). 2 min @ 2000 RPM for PPP concentration.

Companion to the GPS III. Produces fibrinogen-rich platelet-poor plasma concentrate (PPPc) used as a biologic carrier or scaffold-enhancer, often in surgical applications combined with PRP and bone graft.

Terumo BCT / Harvest

4 products
SmartPrep 3 (PRP)
LR-PRP
Spin
Single Spin (automated dual-rotor)
Concentration
~6–7× baseline
Output
PRP kit sizes 7 / 10 / 14 / 20 mL, from ~30 / 60 / 90 / 120 mL blood draws
Centrifuge
Terumo Harvest SmartPrep 3 — proprietary, one-button automated dual spin.

One of the original point-of-care PRP platforms; longstanding orthopedic surgical use. Produces leukocyte-rich PRP with high CD34+ stem cell yields. Frequently used for rotator cuff augmentation, ACL reconstruction adjunct, and chronic tendinopathy.

SmartPrep 3 with Clear PRP
LR-PRP (RBC-depleted)
Spin
Single Spin + RBC filter
Concentration
~6–7× baseline
Output
Same kit sizes as SmartPrep 3 PRP, with ~99% RBC depletion (final ~0.5% hematocrit)
Centrifuge
Terumo Harvest SmartPrep 3 + RBC filter accessory kit.

Accessory kit that removes ~99% of red cells from the SmartPrep PRP product while retaining ~85% of platelets. Reduces inflammation/pain on injection — favored for intra-articular injections where RBC contamination is undesired.

SmartPrep 3 (BMAC)
BMAC
Spin
Automated
Concentration
High MNC and CD34+ concentrate
Output
~7–14 mL BMC from 60 mL bone marrow aspirate
Centrifuge
Terumo Harvest SmartPrep 3 (same device, BMAC program).

Bone marrow aspirate concentrate output from the same SmartPrep platform. Used surgically and clinically for cartilage lesions, AVN, non-union, and recalcitrant tendon/ligament pathology.

AdiPrep (Adipose)
Adipose Concentrate
Spin
Automated on SmartPrep platform
Concentration
Concentrated adipose graft with SVF
Output
~15–30 mL concentrate from 60–120 mL lipoaspirate
Centrifuge
Terumo Harvest SmartPrep 3 (same device, adipose program).

Adipose tissue concentration system on the SmartPrep platform. Used for orthopedic applications (knee OA) and aesthetic/reconstructive fat grafting.

Regen Lab

4 products
RegenKit BCT (BCT-1 / BCT-2 / BCT-3 / BCT-4)
LP-PRP
Spin
Single Spin (thixotropic gel separator)
Concentration
1.7–2.5× baseline (>95% platelet recovery)
Output
4–6 mL PRP from 10 mL blood per tube; multi-tube kits scale to 20–24 mL
Centrifuge
Drucker 642 VFD Plus or Horizon 6 Flex FA (RegenLab-preset). Also compatible with any 45° fixed-angle or 90° swing rotor. Single spin: 1500 × g, 5 min.

A pure, leukocyte-poor PRP using sodium citrate anticoagulant and a thixotropic separating gel. Very low concentration factor but high platelet recovery efficiency. Strong evidence in knee OA and dermatology/aesthetics; popular for its simple single-tube workflow and low patient discomfort.

RegenKit BCT-HF (High Volume)
LP-PRP
Spin
Single Spin
Concentration
~1.7–2.5× baseline
Output
Higher-volume kit (10–30 mL blood per draw) for larger joints or multi-site treatments
Centrifuge
Drucker 642 VFD Plus or Horizon 6 Flex FA. Single spin: 1500 × g, 5 min.

Higher-volume version of the BCT line for larger joint or multi-site treatments.

RegenKit BCT Plus (with ATS)
LP-PRP + Autologous Thrombin Serum
Spin
Single Spin
Concentration
~1.7–2.5× baseline
Output
PRP from 10 mL blood (1 BCT tube) + 1 mL autologous thrombin serum (1 ATS tube) per kit
Centrifuge
Drucker 642 VFD Plus or Horizon 6 Flex FA. Single spin: 1500 × g, 9 min.

Adds an autologous thrombin serum tube to the BCT kit, enabling natural fibrin matrix (PRP gel) formation without exogenous chemical activation. Used for surgical wound healing, ulcers, and tissue scaffolding.

RegenKit A-CP / THT
LP-PRP (MNC-enriched)
Spin
Single Spin
Concentration
Variable
Output
~4.5 mL from 10 mL blood
Centrifuge
Drucker 642 VFD Plus or Horizon 6 Flex FA. Single spin: 1500 × g, 5 min.

Cellular variant of the BCT line, enriching for mononuclear cells (lymphocytes/monocytes) while remaining granulocyte-poor. The THT variant is also distributed by Stryker for joint preservation use.

Apex Biologix

4 products
XCELL PRP
Customizable (LR or LP)
Spin
Double Spin (buffy coat capture)
Concentration
~7–9× baseline (avg 8.93×)
Output
~6.3 mL PRP from 60 mL blood (single spin, dial-driven buffy-coat extraction)
Centrifuge
Drucker Boost 4+ Flex (recommended) or compatible Eppendorf models. Single spin paired with the XCELL Benchtop Processing Station (BPS) for buffy-coat extraction.

Closed buffy-coat-capture system with a benchtop processing station for reproducible separation. Configurable for leukocyte-enriched (tendinopathy) or low-leukocyte (intra-articular OA) preparations. Common in regenerative and interventional orthopedic clinics.

XCELL PRF
Platelet-Rich Fibrin
Spin
Single Spin (no anticoagulant)
Concentration
Fibrin matrix with concentrated platelets
Output
PRF clot or injectable form from 10–20 mL blood (no anticoagulant)
Centrifuge
Drucker Boost 4+ Flex + BPS. Single spin (no anticoagulant).

Platelet-rich fibrin product for aesthetics (microneedling, hair restoration), dental, and orthopedic applications where a sustained-release fibrin scaffold is desired.

XCELL PC (Protein Concentrate)
A2M / Protein Concentrate
Spin
PRP + ultrafiltration
Concentration
High A2M, fibrinogen, anti-inflammatory proteins
Output
~3–14 mL concentrated PPP from 60–120 mL blood (PPP routed through hemoconcentrator)
Centrifuge
Drucker Boost 4+ Flex + BPS + HPH Junior Hemoconcentrator for PPP ultrafiltration.

Concentrates Alpha-2-Macroglobulin (A2M) and related anti-inflammatory proteins from platelet-poor plasma. Used for OA where the protease-inhibitor mechanism of A2M is the therapeutic target.

XCELL BMC
BMAC
Spin
Double Spin
Concentration
Concentrated MNCs + platelets
Output
Variable concentrate from up to 60 mL bone marrow aspirate
Centrifuge
Drucker Boost 4+ Flex + BPS.

Bone marrow aspirate concentrate for cartilage lesions, AVN, and advanced regenerative indications.

Celling Biosciences

4 products
ART PRP
LR-PRP (single chamber)
Spin
Single Spin (dial-driven stack extraction)
Concentration
Variable
Output
3–5 mL from 60 mL blood
Centrifuge
Celling TDZ5B-WS Tabletop Low-Speed Centrifuge — custom-made, exclusively compatible with ART kits. Single spin with dial-driven stack extraction.

Single-chamber thumb-wheel system that lets the operator precisely extract user-defined portions of the centrifuged stack. Used in regenerative orthopedics and surgical adjunct settings.

ART PRP Plus
PRP + A2M / Protein Concentrate
Spin
Single Spin + nano-filtration
Concentration
PRP + concentrated A2M, IRAP, fibrinogen
Output
3–5 mL PRP + 3–5 mL concentrated PPP from 60 mL blood
Centrifuge
Celling TDZ5B-WS Centrifuge + integrated nanopore filter for PPP protein ultrafiltration. Single spin.

Adds an integrated nanopore filter for ultrafiltration of platelet-poor plasma proteins (A2M, IRAP). Targets OA where anti-inflammatory protein concentration matters. Combines PRP injection with a separate concentrated-protein injection from one draw.

ART Two-Step PRP
Customizable (LR or LP)
Spin
Double Spin (dual-chamber)
Concentration
Variable
Output
~3–5 mL PRP from 60 mL blood (operator-defined volume via dial)
Centrifuge
Celling TDZ5B-WS Centrifuge. Double spin (dual-chamber design).

Dual-chamber design lets the operator choose LR-PRP for tendinopathy or LP-PRP for joint indications, with precise dial-controlled layer extraction. Provides flexibility within a single product line.

ART BMC / ART BMC Plus
BMAC
Spin
Single Spin (dial-driven)
Concentration
High-yield MNC concentrate
Output
Operator-defined BMC volume from up to 60 mL bone marrow aspirate
Centrifuge
Celling TDZ5B-WS Centrifuge.

Bone marrow concentrate for cartilage, AVN, non-union, and orthopedic surgical adjunct applications. Plus version adds protein ultrafiltration.

Isto Biologics

3 products
Magellan PRP
Customizable Volume / Concentration
Spin
Automated Dual Spin (optical sensor)
Concentration
Up to ~7× baseline at 6 mL output
Output
3–10 mL PRP per cycle from 30–60 mL blood; up to 3 cycles (90 mL total) per disposable
Centrifuge
Isto Magellan device (formerly Arteriocyte/Medtronic) — proprietary, fully automated dual spin with optical sensors and BioAdaptive technology that auto-adjusts to hematocrit.

Fully automated, closed-loop concentrator using optical sensors and BioAdaptive technology. Operator selects output volume and concentration. Common in hospital ORs (spine, orthopedics, sports surgery) and high-volume clinics. Each disposable can be run 3× per patient.

Magellan MAR0 (BMAC)
BMAC
Spin
Automated Dual Spin
Concentration
Concentrated MSCs + HSCs + platelets
Output
3–10 mL BMC per cycle from 30–60 mL bone marrow aspirate
Centrifuge
Isto Magellan device (same as PRP).

Bone marrow aspirate concentrate output on the Magellan platform. Recovers progenitor cells (MSCs and HSCs), platelets, and growth factors for surgical and clinical regenerative use.

Precise PRP
Customizable (LR or LP)
Spin
Automated on Magellan platform
Concentration
Defined leukocyte fractions
Output
Operator-defined PRP volume from up to 60 mL blood per cycle
Centrifuge
Isto Magellan device (same as PRP).

Newer Isto kit that delivers PRP with precision-controlled cellular fractions — LR-PRP or LP-PRP from the same Magellan device, with defined leukocyte content for evidence-based dosing.

Dr. PRP USA

2 products
Dr. PRP Kit (20 cc)
Customizable (LR or LP)
Spin
Double Spin (closed tube)
Concentration
~6–8× baseline
Output
~4 mL PRP from 18 mL blood
Centrifuge
Compatible with most swinging-bucket lab centrifuges — Biomet GPS, EmCyte Executive, Harvest SmartPrep, ELMI CM-7S commonly cited. Double spin.

FDA 510(k) cleared single-use closed tube with patented one-step separation. >90% platelet recovery, operator-controlled RBC/WBC inclusion. Common in office-based regenerative medicine, aesthetics, hair restoration, and small-joint orthopedic injections. Compatible with most swinging-out centrifuges (Biomet, EmCyte, Harvest, ELMI).

Dr. PRP Kit (30 cc)
Customizable (LR or LP)
Spin
Double Spin (closed tube)
Concentration
~6–8× baseline
Output
~6 mL PRP from 30 mL blood (double spin)
Centrifuge
Same compatibility as 20 cc kit (Biomet, EmCyte, Harvest, ELMI). Double spin.

Higher-volume version of the Dr. PRP kit (510(k) BK251201) for larger joint or multi-site applications. Same closed-tube workflow and visual chamber design.

Stryker

1 product
Stryker RegenKit-THT
LP-PRP (physiologic WBC)
Spin
Single Spin (one-step)
Concentration
Physiologic platelet concentration
Output
~4–5 mL PRP from 8 mL blood per tube (single spin, 8-min cycle)
Centrifuge
Drucker 642 VFD Plus or Horizon 6 Flex FA (RegenLab-preset, Stryker-distributed). Single spin: 1500 × g, ~5 min (8-min cycle).

Stryker Joint Preservation's PRP system, distributed through an exclusive collaboration with RegenLab. One-step, single-tube preparation with an 8-minute centrifuge cycle. Maintains physiologic leukocyte levels (not granulocyte-rich). Used in office-based joint preservation, sports medicine, and surgical adjunct settings.

EXO Advantage

1 product
EXO Advantage Protocol
Customizable (LR or LP)
Spin
Double Spin (AABB-style with precise decant)
Concentration
Operator-defined (dose-focused protocol)
Output
Operator-defined volume (typically 30–60 mL blood input); LR or LP selected at decant
Centrifuge
Miniaturized blood-banking centrifuge — specific device not publicly disclosed by EXO Advantage. AABB-style double spin with precise decant.

A miniaturized adaptation of the AABB/Red Cross blood-banking workflow — two centrifugation steps with a precise decant at the plasma/WBC/RBC interface to transition between LP-PRP and LR-PRP from a single draw. Marketed on cost per billion platelets delivered rather than a fixed concentration factor, with emphasis on hitting dose thresholds from the literature (referencing the 2024 Boffa et al. AJSM paper on platelet concentration in knee OA). EXO Advantage also offers separate fibrin/exosome, BMAC, and adipose protocols on the same platform.

Note: System specifications (platelet concentration factors, leukocyte profiles, output volumes) are drawn from manufacturer literature and peer-reviewed comparative studies. Actual yields vary by patient hematocrit, draw volume, and operator technique. Always confirm current specifications and FDA-cleared indications directly with the manufacturer. PRP system selection should account for the patient's diagnosis, treatment goal (tendon vs. joint vs. surgical adjunct), and the latest evidence on leukocyte-rich vs. leukocyte-poor preparations.